STUDY ON INFLUENCE OF PARASITIC OSCILLATIONS OF VIBRATOR MASS FLOW TRANSDUCER ON MEASUREMENT RESULTS
Abstract
The parasitic oscillations of the vibrator mass flow transducer caused by the force point shift from the axis of symmetry of the U-tube are investigated. The defining formulas of the absolute and adduced additive errors are received. The criteria on reducing these errors in the flowmeter design are developed.
References
1. Tests of various configurations of Coriolis mass flowmeters / Pradeep Gupta, K.Shrinivasan, S.V. Prabhu//Elsevier. ScienceDirect. Measurement. – #39(2006). – S.296-307.
2. Performance evaluation of an indigenously designed copper (U) tube Coriolis mass flow sen-sors/Satish C. Sharma, Pravin P.Patil*, Major Ashish Vasudev, S.C. Jain//Elsevier. ScienceDirect. Meas-urement. – #43 (2010). – S.1165-1172.
3. Strelkov S.P. Vvedenie v teoriyu kolebanij / S.P. Strelkov. – M.: Nauka, 1964. – 440 s. – In Russian.
4. Shkalikov V.S. Izmerenie parametrov vibracii i udara / V.S. Shkalikov, V.S. Pellinecz, E.G. Isakovich i dr. – M.: Izd-vo standartov, 1980. – 278 s. – In Russian.
Review
For citations:
Parkhomenko G.G. STUDY ON INFLUENCE OF PARASITIC OSCILLATIONS OF VIBRATOR MASS FLOW TRANSDUCER ON MEASUREMENT RESULTS. Vestnik of Don State Technical University. 2012;12(2-1):47-53. (In Russ.)